• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem Prolapse (2475)
Event Date 06/07/2021
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic operating console presented problems of higher inflow than the vacuum, which caused eye prolapse during a cataract surgery.The surgeon was able to stabilize the eye and implanted the intra-ocular lens (iol) without further issue.Additional information has been requested.
 
Manufacturer Narrative
The company service representative, examined the system.And was unable to confirm or replicate the reported event.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The system was found to meet specifications.Therefore, the root cause of the reported event cannot be determined conclusively.Data will continue to be monitored for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key12126939
MDR Text Key260218589
Report Number2028159-2021-00716
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517633
UDI-Public00380657517633
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-