Patient had a repeat cesarean section with bilateral tubal ligation at 32 weeks for severe preeclampsia.
After second count, scrub tech noted that the tip of a needle driver was missing.
Mds, circulating nurse, and charge nurse notified immediately.
Surgical team checked around sterile field, but could not find the missing tip.
Patient closed up and brought out to pacu (post anesthesia care unit) for recovery.
Md and anm (assistant nurse manager) at bedside to notify patient and family about the possibility of the missing needle driver tip being retained during surgery and that she will receive an x-ray at bedside to determine if it was retained.
Patient and husband verbalized understanding.
X-ray technician came to bedside, results were read by md.
Md came to bedside to notify patient and support person that nothing was seen on the x-ray.
Patient and husband verbalized understanding.
This device is a consumable and the instrument involved is not maintained by clinical engineering.
No further action required by clinical engineering.
|