• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD TRANS-RAY 7FR. 40CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD TRANS-RAY 7FR. 40CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0514
Device Problem Restricted Flow rate (1248)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2021
Event Type  malfunction  
Manufacturer Narrative
Event site postal code: (b)(6). The device was not returned and could not be evaluated. It was discarded by the user. We are unable to confirm the reported event. If new information becomes available, a supplemental report will be submitted. Complaint record id # (b)(4).
 
Event Description
It was reported that immediately after insertion of the intra-aortic balloon (iab), the console frequently generated check iab catheter and catheter restriction alarms. The iab was checked and no issues were found. The console was swapped out with a different one, but the problem continued. The extension tube was removed and the expansion of the iab was confirmed by fluoroscopy while pushing it by hand with a syringe. The iab was removed and replaced with a new one, but the issue was not resolved and the cause could not be determined. When the augmentation pressure was lowered, the drive became stable and the alarm could be avoided so therapy was continued. Approximately 16 hours later, the second iab was also removed. It was noted that another manufacturer's sheath was used. There was no patient harm or adverse event reported. This report is for the first iab used in this event. A separate report will be submitted for the second iab used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTRANS-RAY 7FR. 40CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
dorota wolpiuk
15 law drive
fairfield, NJ 
MDR Report Key12134951
MDR Text Key260677810
Report Number2248146-2021-00442
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/05/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number0684-00-0514
Device Lot Number3000123658
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/05/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-