• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. AS LVP 20D 3SS CV; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. AS LVP 20D 3SS CV; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 2426-0007
Device Problems Complete Blockage (1094); Restricted Flow rate (1248); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: one sample was received for quality investigation.The customer complaint of tubing kinked was verified by visual inspection.Evaluation of the sample submitted confirmed a kink in the tubing below the drip chamber causing for complete occlusion of the fluid flow.A device history record review could not be performed on model 2426-0007 because a lot number was not provided by the customer.The root cause for this issue is the coiling and packaging of the infusion set before the solvent used to joint the drip chamber to the tubing, is allowed to dry fully.This causes the tubing to be kinked and effecting the ability for the infusion set to flow correctly.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that a kink in the as lvp 20d 3ss cv tubing prevented it from being primed.The following information was provided by the initial reporter: "the nurse thought the kink was just a bend in the tubing from packaging but the tubing is actually blocked and would not prime.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS LVP 20D 3SS CV
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12140227
MDR Text Key261485595
Report Number9616066-2021-51491
Device Sequence Number1
Product Code FPA
UDI-Device Identifier50885403227996
UDI-Public50885403227996
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2426-0007
Device Catalogue Number2426-0007
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-