• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-60
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Through follow-up communication livanova learned that the customer may have accidentally powered off the unit during the procedure.Investigation is ongoing to clarify the cause of the reported event.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) monitor went blank during a procedure.Based on information currently available a pump stop cannot be excluded.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the s5 console read-out analysis has been performed and no errors related to any malfunction of the cp5 device were found.Based on functional test and read-out analysis performed, the most likely root cause is an user error accidentally turning off the device.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12141416
MDR Text Key260722701
Report Number9611109-2021-00363
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-