Model Number 3186
Device Problems Fracture (1260); High impedance (1291); Low impedance (2285)
Patient Problems Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
Related manufacturer reference number: 1627487-2021-15467.It was reported the patient is experiencing high and low impedances on the leads.Surgical intervention may take place at a later date.
 
Event Description
Additional information received indicates the lead was fractured.Surgical intervention took place wherein one lead was explanted and replaced.Therapy was restored.
 
Manufacturer Narrative
E1 - initial reporter.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key12143667
MDR Text Key260747790
Report Number1627487-2021-15468
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05414734401708
UDI-Public05414734401708
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/28/2014
Device Model Number3186
Device Catalogue Number3186
Device Lot Number3710356
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MODEL 3186 LEAD; MODEL 3186 LEAD
Patient Outcome(s) Other;
Patient Weight91
-
-