DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: (10/213).A getinge field service engineer (fse) was dispatched to evaluate this unit and was unable to duplicate reported failure.The fse performed full functional check and safety test and the unit passed all functional and safety test per factory specifications.The iabp was released to the customer and cleared for clinical service.Upon completion of our investigation, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during use on a patient, the fiber optic reading was not communicating to the cs300 intra-aortic balloon pump (iabp).There was no harm or injury to the patient and no adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, e2, e3, g3, g6, g7, h2, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person), h4.
|
|