Model Number 0998-00-3023-53
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: (10/213).A getinge field service engineer (fse) was dispatched to evaluate this unit and was unable to duplicate reported failure.The fse performed full functional check and safety test and the unit passed all functional and safety test per factory specifications.The iabp was released to the customer and cleared for clinical service.Upon completion of our investigation, a supplemental report will be submitted.
 
Event Description
It was reported that during use on a patient, the fiber optic reading was not communicating to the cs300 intra-aortic balloon pump (iabp).There was no harm or injury to the patient and no adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, e2, e3, g3, g6, g7, h2, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person), h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12143866
MDR Text Key264485870
Report Number2249723-2021-01469
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-