Model Number 0998-00-0800-53 |
Device Problems
Filling Problem (1233); Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To inspect the issue the fse removed the chassis cover and inspected the pneumatic tubing, he ensured tubing was connected and not kinked.He then re-ran autofill and which passed and instructed the customer to burn in the system for three days.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter noted is (b)(6).
|
|
Event Description
|
It was reported that during a routine check the cardiosave intra-aortic balloon pump (iabp) had an autofill failure.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: h6 (type of investigation), (investigation conclusions).Corrected fields:h6 (medical device -problem code).Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (jul 2019 through jun 2021) was reviewed.There were no triggers identified for the review period.
|
|
Search Alerts/Recalls
|