• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue. The collar around the ecg jack was loose. The fse tightened the collar. The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) ecg jack was loose. It is unknown the circumstances under which the event occurred. It is also unknown if there was patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12146177
MDR Text Key265573970
Report Number2249723-2021-01481
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/19/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2013
Is the Device Single Use? No
Type of Device Usage Reuse

-
-