BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM US ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number NPFS02000
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2019
Event Type  malfunction  
Event Description
It was reported that, during a ukr surgery, the anspach foot pedal did not connect: it was unplugged and replugged with no success; it was also powered down and up with no success. Both male and female ends of connectors are perfect. The cable was jiggled while connected and anspach foot pedal was recognized. No delay was reported and the surgery finished with the same device. The patient was not injured. The investigation found that the anspach foot pedal was not recognized, it showed as not connected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVIO SURGICAL SYSTEM US
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd. suite 40
plymouth MN
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
suite 100
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12149341
MDR Text Key260991015
Report Number3010266064-2021-00546
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628416
UDI-Public00885556628416
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberNPFS02000
Device Catalogue NumberNPFS02000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 07/12/2021 Patient Sequence Number: 1
Treatment
120013S, SVCE, ANSPACH FOOT CONTROL, M30313234810
-
-