Model Number ROB10013
Device Problems Overheating of Device (1437); Failure to Calibrate (2440)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2021
Event Type  Injury  
Event Description
It was reported that during the set up for a cori assisted tka surgery, the real intelligence robotic drill terminal end was very hot to the touch. When running the kpc test it repeatedly failed random exposure and seemed to sound and run sluggishly. They tried and failed approximately six times. They spoke to the engineers and then did some testing in another room. Once the drill had cooled significantly, it passed the kpc test. They tried multiple times with four different long attachments to confirm and passed every time. The procedure was completed, with a delay (less than 30 minutes) by changing to manual procedure. Patient was not harmed as consequence of this problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREAL INTELLIGENCE ROBOTIC DRILL
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12160344
MDR Text Key261292169
Report Number3010266064-2021-00552
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 07/13/2021 Patient Sequence Number: 1
-
-