It was reported that during the set up for a cori assisted tka surgery, the real intelligence robotic drill terminal end was very hot to the touch.
When running the kpc test it repeatedly failed random exposure and seemed to sound and run sluggishly.
They tried and failed approximately six times.
They spoke to the engineers and then did some testing in another room.
Once the drill had cooled significantly, it passed the kpc test.
They tried multiple times with four different long attachments to confirm and passed every time.
The procedure was completed, with a delay (less than 30 minutes) by changing to manual procedure.
Patient was not harmed as consequence of this problem.
|