CYBERONICS, INC. PROGRAMMING WAND Back to Search Results
Model Number 201
Event Date 09/29/2008
Event Type  Malfunction  
Event Description

Reporter indicated a vns programming wand worked intermittently. The wand was returned to the manufacturer and analysis confirmed the event. The reset switch was stuck in the on state, which prevented device communication.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING WAND
Manufacturer (Section D)
CYBERONICS, INC.
houston TX
Manufacturer Contact
nydia grimes
100 cyberonics blvd.
ste 600
houston , TX 77058
2812287200
MDR Report Key1216052
Report Number1644487-2008-02627
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 09/29/2008
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/29/2008
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number201
Was Device Available For Evaluation? Yes
Date Returned to Manufacturer10/13/2008
Is The Reporter A Health Professional? Yes
Date Manufacturer Received09/29/2008
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured10/01/2001
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-