Model Number 45007
Device Problems Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2021
Event Type  malfunction  
Event Description
It was reported that the jetstream catheter was torn. A jetstream xc catheter, 2. 1mm, was selected for use in an atherectomy procedure in the superficial femoral artery(sfa)/tibial peroneal trunk (tpt). During the procedure, when the device was through the tpt vessel, the outer sheath of the jetstream device tore. This tear caused fluid to spray out of the catheter and it was no longer able to be advanced. The device was removed. The procedure was completed via balloon angioplasty. No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameJETSTREAM XC
Type of DeviceCATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork IRELA ND
EI IRELAND
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12164273
MDR Text Key261451378
Report Number2134265-2021-08869
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0026848468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-