Model Number 71430432
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Event Description
It was reported that, three (3) patella trials gii oval resurf pat trial 35mm, gii oval resurf pat trial 29mm and gii oval resurf pat trial 32mm, have deep scratches and small chunks missing.As this was noticed upon field inspection, there was not patient involvement.
 
Manufacturer Narrative
H6: the device, intended for use in treatment, was returned for evaluation.A visual inspection confirms deep gouges/burrs and scratches.The instrument exhibits significant amount of use and wear.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint, however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OVAL RESURF PAT TRIAL 32MM
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12164317
MDR Text Key261453663
Report Number1020279-2021-05928
Device Sequence Number1
Product Code JWH
UDI-Device Identifier03596010484925
UDI-Public03596010484925
Combination Product (y/n)N
PMA/PMN Number
K150241
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71430432
Device Catalogue Number71430432
Device Lot Number15KM01280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Date Manufacturer Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-