• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Device not returned: at this time, the customer has not requested getinge to evaluate the iabp.Additional information is being requested from the customer with regard to the repair and status of the iabp.A supplemental report will be submitted if this information is provided to us.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had a fiber-optic sensor module failure message.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h6(investigation type & investigation conclusions), h10, h11 corrected fields: d1, g1(contact person), h4.A getinge field service engineer (fse) was not dispatched to evaluate the iabp unit because the chief of perfusion states that their hospital does not use fiberoptic balloon catheters, but is seeing the "module failure" message on the pump.The chief of perfusion states it was seen before on this pump prior to this patient.There are no adverse patient outcomes and the pump is working well currently.The message is constant.The fse advised the chief of perfusion to have this pump looked at for service.The chief of perfusion reports that they have someone in house who services their equipment and that person will be contacted.No return calls or emails received.The fse reported that that the local service rep said iabp can be used with alarm.If additional information is received, we will reopen and update the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12167060
MDR Text Key261533205
Report Number2249723-2021-01521
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-