Catalog Number UNKNOWN
Device Problems Difficult to Remove (1528); Structural Problem (2506); Detachment of Device or Device Component (2907)
Patient Problems Perforation of Vessels (2135); Device Embedded In Tissue or Plaque (3165)
Event Type  Injury  
Manufacturer Narrative
Catalog # is unknown but referred to as celect. Occupation: non-healthcare professional. Investigation: the following allegations have been investigated: fracture (fractured piece remains in patient), vena cava (vc) perforation, embedded, complex removal. The reported allegations have been further investigated based on the information provided to date. Filter fracture has been reported and may be either symptomatic or asymptomatic. Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e. G. , a renal vein). Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e. G. , a surgical or endovascular procedure in the vicinity of a filter). Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported. Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures. Filter interacts with ivc wall, e. G. Penetration/perforation/embedment. This may be either symptomatic or asymptomatic. Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e. G. , a surgical or endovascular procedure in the vicinity of a filter). Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration. Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up. The risks/benefits of filter retrieval should be considered for each patient during follow-up. Once protection from pe is no longer necessary, filter retrieval should be considered. Filter retrieval should be attempted when feasible and clinically indicated. Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e. G. , a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications). For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth. The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set. It may also be retrieved with the (b)(4) vascular retriever. Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques. The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position. The safety or effectiveness of these alternative retrieval techniques has not been established. Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve. For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth. Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to specifications. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava. Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. This report includes information known at this time. A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2010. Approximately 10 years and 3 months after receiving the filter implant, the patient underwent a complex percutaneous retrieval attempt. At least two (2) struts were perforating the ivc, and one (1) fractured limb was embedded in the ivc. After multiple attempts, the filter was removed but the fractured strut remains embedded in the wall of the ivc. Hospital and medical records have been requested, but not yet provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer Contact
lissi walmann
sandet 6
bjaeverskov 4632 
56868686
MDR Report Key12168000
MDR Text Key261822658
Report Number3002808486-2021-01586
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation
Type of Report Initial
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 07/14/2021 Patient Sequence Number: 1
-
-