MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24630
Device Problems Detachment of Device or Device Component (2907); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/14/2021
Event Type  malfunction  
Event Description
It was reported that a blade detachment occurred. The 70% stenosed target lesion was located in the non tortuous and non calcified brachial vein. A 5. 00mm/2. 0cm/90cm peripheral cutting balloon was selected for use. During the procedure, resistance was felt when the catheter was tried to insert into the sheath. The device was removed via the sheath and it was noted that one of the blades was detached. The procedure was completed with another of same device. No patient complications were reported and patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2CM PERIPHERAL CUTTING BALLOON
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12168030
MDR Text Key261537804
Report Number2134265-2021-08721
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeSN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/05/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number24630
Device Catalogue Number24630
Device Lot Number0026311117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-