Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the iabp, but was unable to reproduce the reported "fo module failure" issue.However, the fse was able to verify the alarms "fos continuous bit failed" in the logs, but no other fiber optic failures noted in the fault logs.During the fiber optic testing, found defective ibp adapter cable, to fix the issue the fse replaced the ibp adapter cable and completed the fiber optic test successfully.He then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full event site name is (b)(6) medical center.
|