INTEGRA LIFESCIENCES SWITZERLAND SAR CERTAS INLIN VLV SPHN/UNIT CAT; CERTAS PLUS W/ SG Back to Search Results
Catalog Number 828806
Device Problem Obstruction of Flow (2423)
Patient Problem Failure of Implant (1924)
Event Date 06/24/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported that the codman certas valve with sg was implanted to a (b)(6) female patient via v-p shunt on an unknown date with an unknown setting.Obstruction was suspected.Therefore, the valve was removed and replaced on (b)(6) 2021.No further information was provided.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields:  d4, d9, g3, g6, h2, h3, h6, h10 the certas valve was returned for evaluation.Failure analysis - the position of the cam when valve was received was at setting 4.The valve was visually inspected; the needle guard was slightly raised, and needle holes in the needle chamber were noted.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.The needle chamber was dismantled; the needle guard was bent, and glue traces were noted on the needle guard and the silicone base.Root cause - at the time of the investigation no occlusion was noted with the valve.The possible root cause for the issue reported by the customer is probably due to wrong handling as noted in the "ifu": do not fold or bend the valve, folding or bending might cause rupture of the silicone housing, needle guard disc dislodgement or occlusion of the fluid pathway.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAS INLIN VLV SPHN/UNIT CAT
Type of Device
CERTAS PLUS W/ SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
MDR Report Key12170745
MDR Text Key261678260
Report Number3013886523-2021-00307
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K143111
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number828806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2021
Date Manufacturer Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age64 YR
-
-