Catalog Number 500-BAS-UK-10
Device Problems Impedance Problem (2950); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2021
Event Type  malfunction  
Manufacturer Narrative
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event Description
As patient impedance increases, the sensing of a child patient when an adult pad-pak is inserted will likely result in a reduced shock energy being delivered to the patient which may result in adverse consequences.No patient involvement.
 
Manufacturer Narrative
Heartsine's investigation confirmed the reported fault as the device was found to be giving incorrect child patient prompts with an adult pad-pak installed.The investigation determined a failure of the reed switch which controls whether the device recognises an adult or child patient pad-pak insertion.If an child patient is detected but an adult is connected, the higher the impedance the lower the shock energy.There is potential to cause an adverse event but it is dependent on the impedance of the patient.
 
Event Description
As patient impedance increases, the sensing of a child patient when an adult pad-pak is inserted will likely result in a reduced shock energy being delivered to the patient which may result in adverse consequences.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,500P,PP03,INTL EN,500-BAS-UK-10
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
MDR Report Key12171567
MDR Text Key263539781
Report Number3004123209-2021-00183
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number500-BAS-UK-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Date Manufacturer Received08/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-