MPRI ATTAIN OVER THE WIRE LEAD; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number 4193
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Undesired Nerve Stimulation (1980); Device Overstimulation of Tissue (1991); Perforation of Vessels (2135); Discomfort (2330)
Event Date 06/22/2021
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: n161 crt-d implanted, (b)(6) 2012.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was experiencing diaphragmatic and phrenic nerve stimulation from the left ventricular (lv) lead which was causing discomfort.A replacement procedure was scheduled.After the lead was removed, the patient became hypotensive due to a tear in the coronary sinus.A pericardiocentesis was performed and a sternotomy was completed to repair the tear.A new lead will be implanted after the patient heals.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the partial lead was returned in segments, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN OVER THE WIRE LEAD
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key12172222
MDR Text Key261656717
Report Number2649622-2021-14243
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/09/2005
Device Model Number4193
Device Catalogue Number4193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Date Manufacturer Received09/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6944-65 LEAD, 4470 LEAD; 6944-65 LEAD, 4470 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age51 YR
-
-