• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP MRI 6CF INT W SP, ATT, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP MRI 6CF INT W SP, ATT, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 8806060
Device Problems Unsealed Device Packaging (1444); Peeled/Delaminated (1454); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.The catalog number identified has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified.(expiry date: 09/2022).
 
Event Description
It was reported that prior to the placement of power port, the outer packaging was allegedly torn.There was no patient contact.
 
Event Description
It was reported that prior to the placement of power port, the outer packaging was allegedly torn.There was no patient contact.
 
Manufacturer Narrative
H10: the initial mdr was inadvertently submitted with a g3 date of 06/18/2021.The correct g3 date is 06/16/2021.H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported breach of sterile barrier as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiry date: 09/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Event Description
It was reported that prior to the placement of power port, the outer packaging was allegedly torn.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one unsealed powerport isp mri implantable port kit was returned for evaluation.The investigation is confirmed for the identified delaminated issue as the tyvek lid was noted to be delaminated into two layers.The investigation is also confirmed for the identified unsealed device packaging issue as seal transfer was noted along the lip of the tray and small layer of the outer tyvek lid was noted on the tray corner.Further during visual evaluation, the outer tyvek lid was noted to be partially peeled.However, the investigation is inconclusive for the reported torn package issue as the exact circumstances at the time of opening the package is unknown and the package was returned with partially peeled tyvek lid.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2022), g3, h6 (device).H11: h6 (method, result).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP MRI 6CF INT W SP, ATT, SL
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12173447
MDR Text Key261710408
Report Number3006260740-2021-02757
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8806060
Device Lot NumberREEX0573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-