• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2021
Event Type  malfunction  
Manufacturer Narrative
The reported problem was resolved through troubleshooting with olympus technical support via the phone.Upon troubleshooting the issue, it was determined that the red, green, blue (rgb) video cable, from the cv-180 to the monitor, was crimped.To resolve the problem, the reporter ordered a new rgb cable.The investigation is ongoing and a definitive root cause of the reported event has not been determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
A user facility reported to olympus that the image displayed was "shaking." as reported to olympus, the problem was identified during inspection and there was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, there was no malfunction in the device concerned and that the image was shaking because there was an abnormality in the rgb video cable.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12174261
MDR Text Key267529447
Report Number8010047-2021-08907
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-