CORDIS CORPORATION TRAPEASE PVCF FEM/JUG 55CM CSI; FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number 466P306AU |
Device Problem
Unintended Movement (3026)
|
Patient Problems
Perforation (2001); Perforation of Vessels (2135)
|
Event Date 02/24/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
It was reported that a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused filter perforation and tilt.The indication for the filter implant has not been provided and there is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease ivc filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Ivc filter tilt has been associated with practitioner technique and vessel anatomy, specifically asymmetry and tortuosity.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.The timing and mechanism of the tilt and perforation has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event.Without procedural films or post implant imaging available for review, the reported filter tilt and perforation could not be confirmed or further clarified.There is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Event Description
|
As reported by the legal department, the patient underwent placement of the trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, filter perforation, tilt.As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
|
|
Event Description
|
Additional information received per the medical records state that the indication for filter placement was recurrent pulmonary emboli and noncompliance with anticoagulation therapy.The filter was deployed via the patient's right internal jugular vein.The filter was placed just below the level of the renal veins inflow.After instillation of the filter, a central venous line was placed at the superior vena cava-right atrial (svc-ra) junction.The catheter was sutured in place.The patient tolerated the procedures well. additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter struts outside the inferior vena cava (ivc) and filter tilt.The patient became aware of the reported events approximately eleven years and three months after the index procedure.The patient also experienced shortness of breath and chest pain.
|
|
Manufacturer Narrative
|
After further review of additional information received, the following sections have been updated accordingly: a2, b3, b4, b5, b7, d1, d4, d10, g2, g3, g6, h1, h2, h4 and h6.As reported, the patient underwent placement of a trapease vena cava filter.The indication for the filter placement was reported as recurrent pulmonary embolism (pe) in the setting of non-compliance with anticoagulation therapy.The filter was implanted via the right internal jugular vein and placed in an infrarenal position.The patient is reported to have tolerated the procedure well.More than eleven years after the filter implantation, the patient became aware that the filter had tilted, and filter struts had perforated outside the inferior vena cava (ivc).The patient further reported having experienced shortness of breath and chest pain associated with the filter.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The optease retrievable vena cava filter is indicated for use in the prevention of recurrent pe via percutaneous placement in the ivc for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt and ivc perforation events could not be confirmed and the exact cause could not be determined.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Additionally, the timing and mechanism of the filter tilt is unknown.It is unknown if the tilt contributed to the reported perforation.A review of the instructions for use (ifu) notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.Due to the nature of the complaint, the reported shortness of breath and chest pain experienced by the patient could not be confirmed and the exact cause could not be determined.These clinical events do not represent evidence of a device malfunction.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Search Alerts/Recalls
|
|
|