• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. REDIGUARD IAB: 7FR 30CC SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. REDIGUARD IAB: 7FR 30CC SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000262
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
It was reported that "on 14th jun, after pci surgery, iab supported, the guide wire puncture was difficult in the process of catheter placement, and the machine could not counterpulsate normally after balloon insertion, accompanied by machine alarm, no pressure counterpulsation waveform. After replacement of new catheter, the alarm was restored, and the machine counterpulsation was normal. " there was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREDIGUARD IAB: 7FR 30CC
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12180559
MDR Text Key261936358
Report Number3010532612-2021-00216
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K981660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/31/2021
Device Model NumberIPN000262
Device Catalogue NumberIAB-S730C
Device Lot Number18F20A0008
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-