Model Number 0684-00-0479-01 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
|
|
Event Description
|
It was reported that after insertion of the intra-aortic balloon (iab), the console was unable to pump.The iab was replaced with a new one and therapy was continued successfully.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
Occupation: perfusionist.The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Manufacturer Narrative
|
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id (b)(4).
|
|
Event Description
|
N/a.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter.The sheath and step dilator were also returned.The inner lumen was found occluded with dried blood.The occlusion was able to be cleared.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The iab was placed on the cs300 pump and the iab fully inflated.No alarm sounded from the pump.The reported event cannot be confirmed by the evaluation.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period aug-19 through jul-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #: (b)(4).
|
|