Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit and was able to confirmed batteries not charging.No charging voltage was present.The fse placed order for repair parts.When the fse returned to the site on 6/29/2021, the batteries (0146-00-0097) , the power supply assy (0997-00-1180) and power supply monitor board (0670-00-1166 ) were replaced.This corrected the issue.The fse stated that batteries were weak and completely depleted.Device passed all functional and safety tests according to factory specifications.The iabp was returned to customer and cleared for clinical use.Upon completion of our investigation, a supplemental report will be submitted.
 
Event Description
It was reported that during routine check, the battery in the cardiosave intra-aortic balloon pump (iabp) was dead.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Analysis of production: (3331/213) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period jul 2019 through jun 2021 was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key12182285
MDR Text Key261995412
Report Number2249723-2021-01553
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/20/2021
Removal/Correction NumberN/A