• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SPECTRA WAVEWRITER; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SPECTRA WAVEWRITER; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1160
Device Problems Overheating of Device (1437); Positioning Problem (3009); Migration (4003)
Patient Problems Burning Sensation (2146); Superficial (First Degree) Burn (2685); Skin Inflammation/ Irritation (4545)
Event Date 01/22/2021
Event Type  Injury  
Manufacturer Narrative
Exact date unknown, event occurred few months from the date the manufacturer became aware of the event.
 
Event Description
It was reported that the patient was experiencing burning with signs of redness at the ipg site when charging.The patient described the feeling as burning from the inside.It was also believed that the patients ipg had migrated.Database analysis revealed charge profile shows signs of poor charger to ipg coupling and the charger thermistor triggering often which can delay charging times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA WAVEWRITER
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12185263
MDR Text Key262158555
Report Number3006630150-2021-03818
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729951254
UDI-Public08714729951254
Combination Product (y/n)N
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/22/2020
Device Model NumberSC-1160
Device Catalogue NumberSC-1160
Device Lot Number344742
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
-
-