Model Number 3662 |
Device Problems
Wireless Communication Problem (3283); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 06/30/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Event Description
|
It was reported the patient's ipg would not communicate with their controller.Surgical intervention may be undertaken to address the issue.
|
|
Manufacturer Narrative
|
A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|
Event Description
|
Additional information received indicates that patient's ipg would not communicate with external devices.Reportedly, patient underwent an unrelated surgery wherein the ipg was not placed into surgery mode.Manufacturer's representative was unable to recover the ipg using clinician programmer.In turn, surgical intervention was undertaken wherein the ipg was explanted and replaced.This addressed the issue.
|
|
Manufacturer Narrative
|
Medical device problem code should have been 3283 - wireless communication problem rather than 4031 - application program freezes, becomes nonfunctional.
|
|