BLUE BELT TECHNOLOGIES NAVIO HANDPIECE THUMBERSCREW; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR100026
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/09/2021
Event Type  malfunction  
Event Description
It was reported that, during machine set up for a navio-assisted surgery, it was found that the navio bone pin 4.0mmx152mm ((b)(4)), the navio bone pin 4.0mm x 127mm ((b)(4)) and the navio handpiece thumberscrew ((b)(4)) had lost the threadings.No delays were reported.It is unknown how the procedure was completed.Patient outcome is unknown.
 
Manufacturer Narrative
H3, h6:the navio handpiece thumberscrew, part number pfsr100026, serial unk and used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation, or field service report.A complaint history review for similar reported/confirmed complaints has identified prior events.While the reported problem was not confirmed during evaluation, the most probable contributing factor(s) to the reported problem is ineffective method of threading thumbscrew into receiver, leading to damaged threads.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE THUMBERSCREW
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key12203414
MDR Text Key262672372
Report Number3010266064-2021-00561
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556648421
UDI-Public00885556648421
Combination Product (y/n)N
PMA/PMN Number
K160537
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR100026
Device Catalogue NumberPFSR100026
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-