• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4). Investigation summary: the complaint device was received at the service center and evaluated. It was reported that the product sheath is stripped. Per service reports, this complaint can be confirmed. During the service evaluation the following defects were identified: cable sheath cut. Keyboard wires too short. Corroded motor. The motor, motor cable and keyboard were replaced to resolve the issues. After repair, the device was found to be working according to the specifications. The faulty parts was identified as the root cause for the device failure during the service evaluation. Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the customer in (b)(6) that the sheath on the micro tornado hp w handcontrol device was stripped. During in-house engineering evaluation, it was determined that the device had corroded motor. There was no procedure nor patient involvement reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICRO TORNADO HP W HANDCONTROL
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key12204381
MDR Text Key262703598
Report Number1221934-2021-02232
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/20/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown
Removal/Correction NumberN/A

-
-