DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0576-01
Device Problems Off-Label Use (1494); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation. If provided we will send a supplemental report with our additional findings. Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console intermittently generated a leak in iab circuit alarm. The insertion was reported to be axillary, which is not the method described in the device instructions for use. The customer was made aware of the off-label use and continued to troubleshoot as though it were a femoral insertion. It was verified that the customer used the iab fill and start key to reset the alarm each time. The customer reported troubleshooting prior to the call to include making sure there was no blood present in the helium tubing, checking integrity of all connections, and checking for external kinking. There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSENSATION PLUS 8FR. 50CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12211765
MDR Text Key262944622
Report Number2248146-2021-00468
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108605
UDI-Public10607567108605
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/26/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot Number3000136074
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/26/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-