|
Catalog Number NNU8LPT |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 08/2023).
|
|
Event Description
|
It was reported that post catheter placement, the device allegedly had obstruction.There was no reported patient injury.
|
|
Event Description
|
It was reported that post catheter placement, the device allegedly had obstruction.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2023), g3 h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Search Alerts/Recalls
|
|
|