A physician reported that the codman hakim programmable valve with sg was implanted to a (b)(6)-year-old male patient with cerebral hemorrhage and protein c deficiency, via v-p shunt on (b)(6) 2018 with a setting of 150mmh2o.However, obstruction was observed, and it was found that the contrast agent did not flow.Therefore, it was replaced to a new one (hakim) on (b)(6) 2021.The patient has since recovered.
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The hakim valve was returned for evaluation.Device history record (dhr) - the product code 82-3162 with lot 192573 conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected; a needle hole in the needle chamber was noted.The position of the cam when valve was received was 150mmh2o.The valve was hydrated.The valve was leak tested and only leaked from the needle hole.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism, at the time of investigation no functional issues were noted.
|