MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN TRAPEASE FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN TRAPEASE FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466P306X
Device Problems Fracture (1260); Unintended Movement (3026)
Patient Problems Perforation of Vessels (2135); Bowel Perforation (2668)
Event Date 06/30/2021
Event Type  Injury  
Manufacturer Narrative

It was reported that a patient underwent placement of a trapease vena cava filter. The information provided indicated that the filter subsequently malfunctioned and caused ivc filter with 3 mm small bowel, 5 mm aortic, 4mm vertebral, and 5mm mesenteric perforation. Filter is tilted with the apex against the ivc wall. A fractured strut noticed posteriorly perforating the ivc with the tip within the anterior-most margin of the upper endplate of l3 vertebra. The indication for the filter implant has not been provided and there is currently no additional information available for review. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The trapease ivc filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Ivc filter tilt has been associated with practitioner technique and vessel anatomy, specifically asymmetry and tortuosity. Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu). The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters. The ifu also states that filter fracture is a potential complication of vena cava filters. Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut. The timing and mechanism of the reported events has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event. Without procedural films or post implant imaging available for review, the reported filter tilt, fracture and perforation could not be confirmed or further clarified. There is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.

 
Event Description

As reported by the legal brief, the patient underwent placement of an trapease vena cava filter. The filter subsequently malfunctioned and caused injury and damages including, but not limited to: ivc filter with 3 mm small bowel, 5 mm aortic, 4mm vertebral, and 5mm mesenteric perforation. Filter is tilted with the apex against the ivc wall. A fractured strut noticed posteriorly perforating the ivc with the tip within the anterior-most margin of the upper endplate of l3 vertebra.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN TRAPEASE FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key12218737
MDR Text Key263454915
Report Number1016427-2021-05116
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,OTHER
Reporter Occupation
Type of Report Initial
Report Date 07/23/2021
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/23/2021
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number466P306X
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/07/2021
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 07/23/2021 Patient Sequence Number: 1
-
-