Model Number CYF-VHA
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was occurred due to physical stress.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Omsc presumed that the instrument channel port was deformed because the excessive force was applied when attaching or detaching the forceps/irrigation plug.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that during the preparation for use, it was found that the instrument channel port was deformed.There was no report of patient injury associated with the event.The event date was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12223608
MDR Text Key263460259
Report Number8010047-2021-09311
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-