Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot 0010582938 was returned and analyzed. Visual inspection showed the shaft was kinked at three locations: 2. 5 and 8. 5 inches proximal from the tip and 2. 5 inches distal from the strain relief. Deflection functionality of the sheath was within specification. The tip of the sheath was atraumatic and the inner diameter measurement was also within specification. The dilator was received separated from the sheath. Visual inspection of the dilator indicated the dilator was intact without issue. In conclusion, the sheath failed the returned product inspection due to a shaft kink in three locations.
 
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturers investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12229479
MDR Text Key263590457
Report Number3002648230-2021-00362
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/26/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010582938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-