• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-55
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2021
Event Type  malfunction  
Manufacturer Narrative
Device not returned: a supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) had error 53 twice.The iabp was replaced on the patient.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp and in consultation with the mt, the cable winding, safety disk, tital volume disk and 9 v block were exchanged as part of the maintenance.The reported fault # 53 was resolved by replacing the cable winding, safety disk, tital volume disk and 9v block and all functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID INT TYPE E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12231459
MDR Text Key264597400
Report Number2249723-2021-01622
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NA.; UNKNOWN.
Patient SexPrefer Not To Disclose
-
-