Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 double head pump gave a motor control failure error message during priming.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the pump was returned to the manufacturer site for investigation.The analysis of serial read-out of the pump (real time device parameters and setting recording file) was collected and a functional test was performed.The serial read-out confirmed the reported issue and several occurrences of errors associated to resolver were found stored indicating a failure of the resolver converter of the motor control board the reported failure was confirmed also during the functional test.The motor control board was replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Based on all the above facts, a defective motor control board was identified as root cause of the issue.
|
|