• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the implantable cardiac monitor (icm) was suspected to have moved position. The device remains in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREVEAL LINQ
Type of DeviceDETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ 1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12237324
MDR Text Key263861270
Report Number9614453-2021-03010
Device Sequence Number1
Product Code DSI
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial
Report Date 07/28/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/28/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date04/28/2019
Device MODEL NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/15/2021
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured05/03/2018
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-