Model Number 8606500
Device Problems Intermittent Continuity (1121); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still on-going.The results will be provided with a follow-up report.
 
Event Description
It was reported that a ventilator failure occurred during a case.No injury reported.
 
Manufacturer Narrative
The log file was analyzed for the reported date of event, (b)(6) 2021 whereby three motor encoder check errors were found.Thereupon, according to the device safety concept, the system forced a shutdown of the automatic ventilation and posted the corresponding "ventilator failure" alarm.Manual ventilation and monitoring functions remain available in this case.After the event the motor was replaced and was made available for further investigation at manufacturer¿s site.An evaluation of the motor in the laboratory revealed that in several positions the motor did not provide mechanic power due to a mechanically abraded collector disc.In this case the electrical contact between the collector disc and the carbon brushes is partly disrupted.The consequence are fluctuations in the rotating speed so that the measured ventilator piston position does not match with the one calculated by the system's software.To prevent from potentially hazardous output and/or from damages to the ventilator unit the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Finally, it can be concluded that the device has responded to a detected malfunction of a single component as specified.The repair exchange will fully solve the device problem.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that a ventilator failure occurred during a case.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
MDR Report Key12237648
MDR Text Key265867126
Report Number9611500-2021-00321
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8606500
Device Catalogue Number8606500
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-