Model Number N/A
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event Description
It was reported that resistance was felt and the sheath introducer was difficult to insert because the position of the coil wire of the guidewire was misaligned.There was no reported patient injury.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged guidewire is confirmed; however, the exact cause is unknown.One 0.018 in.Stainless steel guidewire in a plastic hoop was returned for evaluation.An initial visual observation showed use residue on the returned sample.The proximal tip of the guidewire was observed to be damaged.A microscopic observation revealed several of the most distal coils of the guidewire were bent and disjointed but connected to the weld tip.Both weld tips were observed to be present and intact.While the exact cause of the damage observed in the returned guidewire is unknown, possible causes include damage during manufacturing, packaging, handling, or use.
 
Event Description
It was reported that resistance was felt and the sheath introducer was difficult to insert because the position of the coil wire of the guidewire was misaligned.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL MICROINTRODUCER KIT PTFE PA INTRODUCER 4.5F X 5CM
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12245726
MDR Text Key264079003
Report Number3006260740-2021-02986
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
PMA/PMN Number
K915167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0668945J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2021
Date Manufacturer Received08/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-