MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 XT PTA CATHETER PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 XT PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number F2020-020
Device Problems Difficult to Remove (1528); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2021
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation. It has not yet been received. A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was performed to treat a lesion in the anterior tibial artery. During advancement of an unspecified command es guide wire thru the lumen of a 2x20mm armada 14 percutaneous transluminal angioplasty (pta) catheter peeling was noticed on the guide wire, which then became stuck, so the devices were removed together. An unspecified armada xt balloon and command es guide wire were used to successfully complete the procedure. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 14 XT PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12245886
MDR Text Key264080917
Report Number2024168-2021-06626
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K121352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberF2020-020
Device Catalogue NumberF2020-020
Device Lot Number00803G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 07/29/2021 Patient Sequence Number: 1
Treatment
COMMAND ES 14
-
-