MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Model Number IPN913776
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported the unit "won't allow adjustment of the humidity level".No patient involvement reported.
 
Manufacturer Narrative
Qn#: (b)(4).The sample was returned for evaluation.A visual exam was performed and the trim ring was noted to be missing and there was a crack on the right back mounting bracket.No other defects were observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where a water reservoir bottle, an adult breathing circuit (880-36kit), a 382-10 conchasmart column, and dual temperature probes were connected to the unit.Air flow was supplied to the unit for a real time operational scenario.Per the user manual, "heated wire circuits are required to change rainout settings", therefore the unit was set up for the bench test using heated wire circuits.The rainout settings were able to be adjusted normally.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
It was reported the unit "won't allow adjustment of the humidity level".No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key12245970
MDR Text Key264083937
Report Number3003898360-2021-00629
Device Sequence Number1
Product Code BTT
UDI-Device Identifier14026704646811
UDI-Public14026704646811
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberIPN913776
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Date Manufacturer Received08/12/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-