The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the tip of the received apex pin is completely broken / snapped off during insertion.The deformation of the screw thread clearly indicates that far too much mechanical force had been applied during insertion which finally led to the breakage of the tip.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿avoid surface damage of implants.Contouring or bending of an implant should be avoided where possible, because it may reduce its fatigue strength and can cause failure under load.If contouring is necessary, allowed by design or prescribed by stryker, the physician should avoid sharp bends, reverse bends or bending the device at a screw hole.Such action must be performed with stryker instruments and in accordance with the specified procedures (see operative technique).¿ based on investigation, the root cause was attributed to a user related issue.The failure was caused due to application of high mechanical force during insertion.If any further information is provided, the complaint report will be updated.
|