Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during an initial hip arthroplasty, the locking ring was dislodged from the cup.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified lock ring is not correctly oriented in lock ring groove.Lock ring tab was stuck in groove and not centered in tab slot window.Damage was noted to the lock ring upon arrival.Lock ring was removed for evaluation and shows deformation and damage.No other damage to the device was noted.Complaint sample was evaluated and the reported event was not confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|