Product analysis: the hawkone device returned coiled and connected to the cutter driver inside 3 biohazard bags.No ancillary devices were returned for analysis.Device was decontaminated with cidex opa solution soak and tergazyme soak.Visual inspection shows the drive shaft broken at the proximal end of the housing.The cutter window is damaged/pinched, and the proximal end of the housing is detached from the anchor pockets.The cutter returned inside the housing at approx.0.6cm distal from the anchor pockets.No damage noted to the guidewire lumen, nosecone, or distal end of the housing assembly.If information is provided in the future, a supplemental report will be issued.
|