It was reported that a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused fracture, perforation, migration, perforation into organ, and perforation abutting organ that causes injury and death.The indication for the filter placement, procedural details, cause of death and medical history have not been provided.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.A review of the ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the ifu as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Without procedural films or post implant images for review the reported event(s) could not be confirmed or further clarified.Death was reported as a result of the events, however with the limited information provided an adequate assessment of the relationship of the filter to the event is not possible.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
As reported by the legal department, the patient underwent placement of the trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, malfunction, including fracture, perforation, migration, perforation into organ, and perforation abutting organ that causes injury and death.As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
|