• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problems Mechanical Problem (1384); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2021
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the product no longer turns and heats up.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: cable resistance value out of tolerance limits.Keyboard resistance value out of tolerance limits.Corroded motor.The motor, motor cable and keyboard were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the customer that during an unknown procedure on (b)(6) 2021, it was observed that the micro tornado hp w handcontrol device did not turn and heated up.During in-house engineering evaluation, it was determined that the motor on the device was corroded.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key12253140
MDR Text Key264305881
Report Number1221934-2021-02338
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Other
Type of Report Initial
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Date Manufacturer Received07/29/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-