• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Component Incompatible (1108)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: (10/213) during a preventative maintenance (pm) service, the getinge field service engineer (fse) discovered the unit failed the drive manifold section of the all manifold tests.Initially, the drive manifold was replaced and the system was retested, but the all manifold tests still failed.After further troubleshooting, the scroll compressor was determined to be faulty, to fix the issue it was replaced, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
 
Event Description
It was reported that during scheduled pm performed by a getinge field service engineer (fse), the unit failed the drive manifold section of the all manifold tests.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: b4,e1(site country),g3,g6,h2,h4,h6(type of investigation, investigation findings, component codes, investigation conclusions), h10.It was reported that during scheduled pm performed by a getinge field service engineer (fse), the unit failed the drive manifold section of the all manifold tests.There was no patient involvement or adverse event reported.To resolve the issue a getinge fse replaced the drive manifold and scroll compressor.After the repair, the unit passed all functional and safety tests per factory specifcations.It has been returned to the customer and cleared for clinical use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12258210
MDR Text Key264462543
Report Number2249723-2021-01663
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/08/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-