DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Component Incompatible (1108)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: (10/213) during a preventative maintenance (pm) service, the getinge field service engineer (fse) discovered the unit failed the drive manifold section of the all manifold tests.Initially, the drive manifold was replaced and the system was retested, but the all manifold tests still failed.After further troubleshooting, the scroll compressor was determined to be faulty, to fix the issue it was replaced, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
|
|
Event Description
|
It was reported that during scheduled pm performed by a getinge field service engineer (fse), the unit failed the drive manifold section of the all manifold tests.There was no patient involvement.
|
|
Manufacturer Narrative
|
Updated fields: b4,e1(site country),g3,g6,h2,h4,h6(type of investigation, investigation findings, component codes, investigation conclusions), h10.It was reported that during scheduled pm performed by a getinge field service engineer (fse), the unit failed the drive manifold section of the all manifold tests.There was no patient involvement or adverse event reported.To resolve the issue a getinge fse replaced the drive manifold and scroll compressor.After the repair, the unit passed all functional and safety tests per factory specifcations.It has been returned to the customer and cleared for clinical use.
|
|
Event Description
|
N/a.
|
|